- Company Info
Brand : Thermo Fisher Scientific
Model : Prima PRO, Sentinel PRO and Prima BT
Specifications
* Analyzer: High-Energy Scanning Magnetic Sector
* Area Classification: General Purpose, or ATEX Zone 1, or Class 1 Div 2 (Z-purged) or Class 1 Div 1 (X-purged)
* Dynamic Range: Faraday Detector: 10ppm – 100% (Application Dependent) Dual Faraday/SEM: 10ppb – 100% (Application Dependent) Analysis Time: 1-30 Seconds (Application Dependent)
* Analytical: Precision Typically 0.1% Relative Standard Deviation (24 Hour) for concentrations in the range 1-100%
* Calibration: Interval Typically 3-12 Months (Application Dependent)
* Service: Interval Analyzer 3 years, External Vacuum Pump 6-12 months (Application Dependent)
* Dimensions: 0.97m Width x 0.72m Depth x 1.48m Height, 240 kg approximate weight
* Power: 115 VAC (±5VAC) or 230 VAC (±10VAC) Consumption 2.0-2.5 kVA
* Environmental: Ideally 20°-25°C. Allowable Range between 12°C and 40°C
* Calibration Ports: Maximum 24 with single RMS or 48 with Dual RMS
* Standard Communications:
– 4-Ports (Modbus (serial RS232, RS422 or RS485) and/or OPC), 13 Digital Outputs, 13 Digital Inputs
* Additionally Available Communication Protocols:
– Allen-Bradley (Data Highway, DH Plus, DH485, 1771, DF-1, Ethernet) Daniel Modbus, Echelon Lonworks, GE Fanuc Genius, MicroTrac Arcnet Modicon (Modbus, Modbus Plus, Remote I/O), Profibus-DP, Reliance Automate